Electron Beams Commissioning and Initial Measurements on an Elekta Synergy Platform Linear Accelerator
- Authors
-
-
Abdurraouf. M. Aghila
Department of Physics, Faculty of Education, University of Tripoli, Libya.Author -
Saad. S. Saad
Higher Institute of Medical Sciences and Technology, Abu-Salim, Tripoli, Libya.Author -
Faraj A. Elmasrub
National Cancer Institute of Sabratha, Sabratha, LibyaAuthor
-
- Keywords:
- lekta Synergy Platform, Electron Beam Commissioning, PDD, Beam Profile, Applicator Factor
- Abstract
-
A linear accelerator (Linac) must be commissioned for use in treating cancer patients. This research work aims to analyze part of the electron beam data produced by the Elekta Synergy Platform Linac at the radiotherapy department at Tripoli University Hospital, Tripoli, Libya. In this paper, percentage depth dose (PDD), beam profile and applicators factor were studied at five different electron beam energies (4, 6, 8, 10 and 12 MeV). The relative measurement part of this study was carried out using a PTW MP3-M 3D water scanning system, and the absolute measurement was taken using a plane parallel ionization chamber and a unidose E electrometer. The analysis for these measurements has yielded the following results: a) The Penetrative qualities of all the electron beam energies were within the manufacturer's tolerance limits of ±1%; b) the maximum values of beam flatness of 2.94 and beam symmetry of 1.93 are within the accepted limits of International Electrotechnical Commission criteria; c) A penumbra maximum value of 1.30 was measured at the 4-MeV electron beam energy and using a 14×14 cm2 applicator size. All the obtained parameters were within the permissible limits. Therefore, the electron beams can be safely used for clinical purposes.
- Downloads
-
Download data is not yet available.
- References
-
Acquah, G. F., & Schiestl, B. (2014). Quality assurance and performance evaluation of an elekta linear accelerator: results from over a year QA experience in Ghana. IJSBAR, 14, 356-363.
Arunkumar, T., Supe, S. S., Ravikumar, M., Sathiyan, S., & Ganesh, M. (2010). Electron beam characteristics at extended source-to-surface distances for irregular cut-outs. Journal of Medical Physics/Association of Medical Physicists of India, 35(4), 207.
Brahme, A., & Svensson, H. (1976). Specification of electron beam quality from the central‐axis depth absorbed‐dose distribution. Medical physics, 3(2), 95-102.
Chosdu, R., Hilmy, N., Tobing, R., Kicky, L. T. K., Razzak, M., Kovacs, A., & Miller, A. (1995). Dosimetry measurements during the commissioning of the GJ-2 electron accelera-tor. Radiation Physics and Chemistry, 46(4-6), 1175-1179.
Elekta customer acceptance test. (2007). Elekta Oncology Systems Ltd. Grawely: Document num-ber 4513 370 1870 08
Gerbi, B. J. (2006). Clinical Applications of High-Energy Electrons. In Levitt, S. H., Purdy, J. A. editors. Technical Basis of Radiation Therapy. Medical Radiology (Radiation Oncology), Berlin, Springer.
Gerbi, B. J., Antolak, J. A., Deibel, F. C., Followill, D. S., Herman, M. G., Higgins, P. D., ... & Khan, F. M. (2009). Recommendations for clinical electron beam dosimetry: supplement to the recommendations of Task Group 25. Medical physics, 36(7), 3239-3279.
Glide‐Hurst, C., Bellon, M., Foster, R., Altunbas, C., Speiser, M., Altman, M., ... & Solberg, T. (2013). Commissioning of the Varian TrueBeam linear accelerator: a multi‐institutional study. Medical physics, 40(3), 031719.
Hospital Physicist Association., (1975). A practical guide to selection dosimetry below 5 MeV for Radiotherapy purposes. London: HPA Report No. 13
International Atomic Energy Agency (IAEA). (1987). Absorbed Dose Determination in Photon and Electron Beams: An International Code of Practice for Dosimetry. Technical Report Series No. 277, Vienna, Austria
International Atomic Energy Agency (IAEA). (2000). Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose To Water. Technical Report Series No. 398, Vienna, Austria
International Commission on Radiation Units and Measurements (ICRU). (1993). Prescribing, re-cording, and reporting photon beam therapy. ICRU report No. 50, Washington, DC, USA
International Electrotechnical Commission. (1984). Report 35-I and 35-II, Medical electrical equipment, Medical Electron Accelerator, Section IV: Functional performance characteristics and report, Geneva, Switzerland
International Electrotechnical Commission. (1997). Medical Electrical Equipment: Dosimeters with ionization chambers as used in radiotherapy standard. IEC-60731, Geneva, Switzerland
International Electrotechnical Commission. (2007). Medical electrical equipment, Medical Electron Accelerators: Functional performance characteristics. IEC-60976, Geneva, Switzerland
Jamshidi, A., Kuchnir, F. T., & Reft, C. S. (1987). Characteristic parameters of 6–22 MeV electron beams from a 25‐MeV linear accelerator. Medical physics, 14(2), 282-288.
Khan, F. M. (2010). Physics of Radiation Therapy. 4thedition. Philadelphia, Lippincott Williams & Wilkins, USA.
Khan, F. M. (1991). Clinical electron-beam dosimetry. Report of AAPM Radiation Therapy Com-mittee Task Group 25. Med. Phys., 18, 73-109.
Kirby, T. H., Gastorf, R. J., Hanson, W. F., Berkley, L. W., Gagnon, W. F., Hazle, J. D., & Shalek, R. J. (1985). Electron beam central axis depth dose measurements. Medical physics, 12(3), 357-361.
Kutcher, G. J., Coia, L., Gillin, M., Hanson, W. F., Leibel, S., Morton, R. J., ... & Wingfield, L. (1994). Comprehensive QA for radiation oncology: report of AAPM radiation therapy committee task group 40. MEDICAL PHYSICS-LANCASTER PA-, 21, 581-581.
Narayanasamy, G., Saenz, D., Cruz, W., Ha, C. S., Papanikolaou, N., & Stathakis, S. (2016). Com-missioning an Elekta Versa HD linear accelerator. Journal of applied clinical medical phys-ics, 17(1), 179-191.
Podgorsak, E. B. (2003). Review of radiation oncology physics: a handbook for teachers and stu-dents. Vienna, Austria: IAE Agency, 19, 133.
Sahool, S. K., Rath, A. K., Mukharjee, R. N., & Mallick, B. (2012). Commissioning of a Modern LINAC for Clinical Treatment and Material Research. International Journal of Trends in In-terdisciplinary Studies, 1(10).
Varatharaj C, Shwetha B, Rekha Reddy B, Ravikumar M, S. Sathiyan, Ganesh K. M. (2016). Phys-ical characteristics of photon and electron beams from a radiotherapy accelerator. Interna-tional Journal of Medical Research and Review Res Rev .4 (7), 1178- 1188.
- Downloads
- Published
- 2023-12-31
- Issue
- Vol. 38 No. 4 (2023)
- Section
- Articles
- License
-
Copyright (c) 2023 Abdurraouf. M. Aghila, Saad. S. Saad, Faraj A. Elmasrub (Author)

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Copyright of the articles Published by Almukhtar Journal of Science (MJSc) is retained by the author(s), who grant MJSc a license to publish the article. Authors also grant any third party the right to use the article freely as long as its integrity is maintained and its original authors and cite MJSc as the original publisher. Also, they accept the article remains published by the MJSc website (except in the occasion of a retraction of the article).
How to Cite
Similar Articles
- Khaled S. Alsalhen, Omalsaad E. I. Omar, Adverse Effects of Organochlorine Pesticide Residues on Biochemical Parameters and Oxidative Stress in Libyan Agricultural Workers , Al-Mukhtar Journal of Sciences: Vol. 37 No. 3 (2022)
- Aeyman M. Hassan, The Design and Implementation of a Libyan Salary Mobile Application using Flutter , Al-Mukhtar Journal of Sciences: Vol. 37 No. 3 (2022)
- Ahmed Y. Habel, Murad M. Aburas, Sery F. Mohamed, Effect of climate factor on some physical soil properties at Al-Jabal Alakhdar, Libya , Al-Mukhtar Journal of Sciences: Vol. 34 No. 3 (2019)
- Haleemah A. Al sheeyab, Hussain Alkhamry, Salem Omaer, Riyad Bendardaf, Cochlear Implant Complications in Children: Libyan Scenario and Experience , Al-Mukhtar Journal of Sciences: Vol. 39 No. 1 (2024)
- Gadalla A. Alhassan, Abdulkader Maleh, Gazzalla I FEDEEL, Distribution of microorganisms in soil horizon at different depths of in Al Jabal Al Akhdar , Al-Mukhtar Journal of Sciences: Vol. 1 No. 1 (1992)
- Muneer A. Abdalla, Seismic Attributes Aided Characterization of Margin Backstepping and Advance along Isolated Carbonate Sequences, Sirt Basin, Libya , Al-Mukhtar Journal of Sciences: Vol. 36 No. 4 (2021)
- Ahmed Rafiq Binamer, Adib Dawud Sulayman Kharufuh, Characteristics of Barbary Sheep Wool as Influeced by Some Factor , Al-Mukhtar Journal of Sciences: Vol. 2 No. 1 (1995)
- Mohammad El-Sayed El-Mor, Biological Studies on Juveniles Fish of Sardinella Aurita (Valenciennes, 1847) in Benghazi Coast on the Meditterranean Sea, , Al-Mukhtar Journal of Sciences: Vol. 20 No. 1 (2008)
- Walid Said Kamel Gaddafi, Gomaa Abdel Salam Afmahieh, The causes and the proportion of extinction of the deer which are living in Libya Studied case of dorcas gazella in south region at jebel alakhder , Al-Mukhtar Journal of Sciences: Vol. 27 No. 1 (2012)
- Al-Hashmi Ali Agleyo, Scanning electron microscopic study of antennal of the female Bee louse fly Braula coeca Nitzsch (Diptera: Braulidae) , Al-Mukhtar Journal of Sciences: Vol. 29 No. 1 (2014)
You may also start an advanced similarity search for this article.